Soft Magnetic Materials DC Measuring Equipment
DX-2012SD soft magnetic material DC measuring system is a comprehensive soft magnetic measuring device capable of automatically measuring the basic magnetization curve and basic hysteresis loop of soft magnetic materials under DC conditions; Which can accurately measure static magnetic characteristic parameters: start Magnetic permeability μi, μm, Bs, Br, Hc, Pu, etc.
Applicable Standards of DX-2012SD hysteresis curves test system
-
GB/T 13012-2008 Measurement method of DC magnetic properties of soft magnetic materials
-
GJB 937-90 weak magnetic material permeability Measurement method
-
GJB 937-90 weak magnetic material permeability Measurement method
-
SJ/T 10281-91 Magnetic parts effective parameters Calculation
-
IEC 60404-4 DC magnetic properties of soft magnetic materials Measurement method
-
IEC 404-7 open circuit magnetic material coercivity Measurement method
Hardware Features of DX-2012SD hysteresis curves test system
-
Automatically complete the demagnetization of closed-circuit samples, the demagnetization method is AC low-frequency demagnetization, with a frequency not higher than 10Hz, to ensure the repeatability of the initial permeability of the metallic soft magnetic material;
-
Using the impact method to test the basic Magnetization curve and permeability curve;
-
Use the impact method to test the static hysteresis of soft magnetic materials Return line and related parameters;
-
Allows the use of scanning method in the open solenoid and love Used in Perstein square circle;
Software Features of DX-2012SD hysteresis curves test system
-
Automatically complete the data calculation and automatically correct the span coefficient. The system supports mainstream operating systems such as WinXP/NT/7/8.
-
Show I(t), B(t) samples Waveform and B(H) hysteresis loop.
-
Can display B(H) hysteresis loop cluster , B(H) magnetization curve, μ(H) permeability curve.
-
The instrument is easy to operate and can be interrupted by human intervention.
-
Has powerful file management functions. It has the functions of automatically saving data, deleting data, and clearing all data.
-
The data file contains complete sampling data and sample parameters And instrument parameters, the file adopts text format, which can be easily input into other software.
-
The test report contains complete graphs, test results, Test conditions and sample parameters.
-
The file system adopts a database format, which can be directly printed or output The test results are transferred to an Excel table.
-
Can be in Chinese and English format, which is convenient for adding users' logos and company names.
-
Has a print preview function for easy adjustment and testing The size and margins of the report.
-
Can print the test report directly, or generate the test report JPG picture file.
-
Can send JPG image files directly via E-mail, Or save the JPG picture file to disk.
-
Supports various types of printers, test reports and printers The paper matches exactly.

Technical Data of DC Hysteresis Graph (DX-2012SD)
Excitation current source
Data acquisition module
AD section
-
Input voltage:±5 volts.
-
AD converter: 16-bit AD, speed 2000-20000 channels/sec.
-
Channel input impedance: 1 Mohm.
-
System accuracy:±0.1% FSR.
-
Resolution: 16 bits.
-
Noise (peak): less than ±1LSB (typical, 500 sampling points)
-
Typical AD conversion speed:
-
USB1.1 interface: 200 calls/sec. Each call can collect 1-16 channels.
-
USB2.0 interface (scan mode): single channel 2000-4000 channels/sec, 16 channel acquisition time 1ms (equivalent to AD conversion speed 16KHz).
-
USB2.0 interface (single-channel average mode): 100 times average, sampling period 2.5ms, equivalent AD conversion speed: 40KHz.
DA part
-
Two-way 16-bit DA converter.
-
Output voltage:±5 volts.
-
Output voltage accuracy: ±0.1%, resolution 16 bits.
-
Setup time of output voltage: less than 100 microseconds.
-
DA output speed: 2000-4000 times/sec (USB2.0 interface)
Electronic integrator
-
Power requirements:±12V.
-
Magnetic flux range: 0.25~0.5~1~2~2.5~5~10~20mwb.
-
Resolution: 0.0001mWb.
-
Accuracy: 0.5%.
-
Drift: 0.001mWb/min.