HALL EFFECT MEASUREMENT SYSTEM

DX-3000 Hall Effect Device

DX-3000 Hall effect measurement system with Keithley instrumentation, configurable magnet options and fully automated software for precise transport property characterisation.

Main Configuration

U.S. Keithley equipment configuration for different resistance ranges.

System (resistance range)2400622027002182A648570017709701271526514488B
Standard (10 mΩ – 10 MΩ)
1111
High sensitivity (0.1 mΩ – 10 MΩ)
11111
High voltage & impedance (0.1 mΩ – 100 GΩ)
11112211

Magnetic parts

Magnetic Field RangeMagnetWater cooling equipmentMagnetic Field Measuring DevicePowerSoftware
0–1T @ 10mmDX-100DX-160 Gaussmeter (accuracy 0.2%)DX-2030 1 kWFully automatic data processing and plotting
0–2.4T @ 10mmDX-130DX-180 Gaussmeter (accuracy 0.05%)DX-10030 3 kWFully automatic data processing and plotting
0–2.6T @ 10mmDX-1788 kW water coolingDX-180 Gaussmeter (accuracy 0.05%)DX-10050 5 kWFully automatic data processing and plotting

Low temperature portion (optional)

Temperature RangeCooling bodyTemperature control device
80–500KLiquid nitrogen thermostatTC202
4–500KCycle refrigeratorTC202
4–325KCryogenic low temperature systemTC202
45–325KCTI small refrigeratorTC202

Sample size

  • Small sample card: 12 × 12 mm².
  • Large sample card: 50 × 50 mm².
  • Probe sample card: 1 × 1 mm² to 30 × 30 mm².

Sample measurement method

  • Hall effect samples such as Si and GaAs using dedicated Hall sample cards.
  • Resistance and I–V samples using 4‑wire or 6‑wire measurement methods for low contact resistance and high accuracy.
  • Support for small, large and probe-style sample cards for different geometries.

Magnetic field environment

Model indexDX-100DX-130DX-180
Pole diameter60 mm (optional)100 mm (optional)120 mm (optional)
Air gapRoom temp.: 10 mm Liquid nitrogen thermostat: 24 mm (optional) ARS refrigerator: 40 mm (optional)Room temp.: 10 mm Liquid nitrogen thermostat: 24 mm (optional) ARS refrigerator: 40 mm (optional)Room temp.: 10 mm Liquid nitrogen thermostat: 24 mm (optional) ARS refrigerator: 40 mm (optional)
Max. magnetic field1.5 T / 0.9 T / 0.6 T2.5 T / 1.5 T / 1.0 T2.7 T / 2.0 T / 1.4 T
Uniformity±1%±1%±1%

Core Functions

  • Measurement of Hall effect and magnetoresistance.
  • Acquisition and plotting of I–V curves at different temperatures and magnetic fields.
  • Acquisition and plotting of R–T curves at fixed magnetic field for resistance versus temperature characterisation.
  • Acquisition and plotting of R–H curves at fixed temperature for resistance versus magnetic field characterisation.

Parameters Measured

  • Sheet resistance and bulk resistivity.
  • Hall coefficient and conductivity type.
  • Hall mobility and carrier concentration.
  • Magnetoresistance parameters including R, ΔR, (ΔR / R)%, (ΔR / R₀)%.

Sample Measurement Methods

  • Hall effect samples such as Si and GaAs using dedicated Hall sample cards.
  • Resistance and I–V samples using 4‑wire or 6‑wire measurement methods for low contact resistance and high accuracy.
  • Support for small, large and probe-style sample cards for different geometries.

Supported Sample Sizes

  • Small sample card: 12 × 12 mm².
  • Large sample card: 50 × 50 mm².
  • Probe sample card: 1 × 1 mm² to 30 × 30 mm².